vol.25 número4Toxoplasmosis aguda en embarazadas asintomáticas de Rio de Janeiro, BrasilCaracterísticas clínicas y anatomo-patológicas de las pacientes que ingresaron al Programa de Control de Tuberculosis con diagnóstico de tuberculosis de mama: Clinical and pathological features of patients treated at the tuberculosis control program with the diagnosis of breast tuberculosis índice de autoresíndice de materiabúsqueda de artículos
Home Pagelista alfabética de revistas  

Servicios Personalizados

Revista

Articulo

Indicadores

  • No hay articulos citadosCitado por SciELO

Links relacionados

  • No hay articulos similaresSimilares en SciELO

Compartir


Revista Medica Herediana

versión impresa ISSN 1018-130X

Resumen

SAMAME, Luz María  y  SAMALVIDES, Frine. Efficacy of the cleaning and disinfection processes of endoscopes in a level III hospital. Rev Med Hered [online]. 2014, vol.25, n.4, pp.208-214. ISSN 1018-130X.

Objectives: To determine the efficacy of the cleaning and disinfection processes of endoscopes in a level III hospital, and to determine the most common pathogens found before and after these procedures. Methods: Case series from August to October 2010. A total of 50 cycles of cleaning-disinfection procedures were evaluated. Culture media (Brilliant blue agar, xylose-lisine-deoxycholate, sulfite bismute, Mc Conkey) as well as coagulase and oxidase tests were used. Results: The mean count of bacteria before the procedures was 835.3 ± 1,114.6 UFC/ml; the median count was 233 UFC/ml; respective values after the procedures were 236.3 ± 700.7 UFC/ml and 10 UFC/ml, respectively, a statistical difference was found (p=0.000001). Bacterial load was positive before the procedures in 88%, and 26% after them. A statistical significant difference was found for Pseudomonas aeruginosa (p = 0.006) and Salmonella entérica (p=0.00001). Bacterial load was positive in 55% after 10 days of using the disinfectant and it was 19% after the ninth day. Conclusions: The cleaning-disinfection process is not effective. Salmonella enterica, Pseudomonas aeruginosa and Escherichia coli were the most frequent isolated pathogens. The high level disinfectant (2% glutaraldehyde) is not effective after the tenth day.

Palabras clave : Gastrointestinal endoscopes; bacterial infections; disinfection; contamination of equipment.