SciELO - Scientific Electronic Library Online

 
vol.84 número3Determinación de transgenicidad y verificación en el etiquetado de alimentos industrializados de maíz en centros de expendio de Lima metropolitanaActividad citotóxica y antiproliferativa de un extracto rico en fucoidan de Lessonia trabeculata sobre células de adenocarcinoma mamario murino 4T1 índice de autoresíndice de assuntospesquisa de artigos
Home Pagelista alfabética de periódicos  

Serviços Personalizados

Journal

Artigo

Indicadores

  • Não possue artigos citadosCitado por SciELO

Links relacionados

  • Não possue artigos similaresSimilares em SciELO

Compartilhar


Anales de la Facultad de Medicina

versão impressa ISSN 1025-5583

Resumo

SALDANA-OREJON, Italo Moisés et al. Interference generated by lipemia in the measurement of biochemical constituents. An. Fac. med. [online]. 2023, vol.84, n.3, pp.286-294.  Epub 30-Set-2023. ISSN 1025-5583.  http://dx.doi.org/10.15381/anales.v84i3.25406.

Introduction.

Turbidity due to lipemia in diagnostic samples is one of the main causes of the appearance of clinically significant biases in the measurement of biochemical magnitudes.

Objective.

To assess the interference by lipemia in the measurement of 25 biochemical constituents in two analyzers with dry chemistry technology (Vitros 7600®) and liquid chemistry (Atellica® Solution).

Methods.

Pre-experimental study with pre and post test. Increasing amounts of a parenteral nutrition lipid emulsion were added to seven aliquots of pooled sera and the influence of the interferent on 25 constituents was determined in duplicate. The relative percentage deviation of the concentration of the constituent due to the influence of turbidity with respect to a sample without interference, was calculated. Tolerance limits for interference were established using three criteria: reagent distributor, desirable systematic error, and maximum permissible error.

Results.

The constituents that presented the greatest biases for the liquid chemistry analyzer were: Phosphorus (-84.72%), ALT (+81.25%) and AST (-75.76%), while for the dry chemistry platform the constituents, ALT (-79.41%), CK (-28.92%) and lipase (+24.85%). Significant interference was detected in a different number of constituents according to the tolerable limit criteria used.

Conclusions.

The different results found according to the methodology and the analyzer used, in addition to the lack of replicability of the tests for the evaluation of interference by lipemia, originates the need to harmonize the processes and establish identical limits of tolerable interference between the laboratories and suppliers of inputs.

Palavras-chave : Bias; Hyperlipidemias; Diagnostic Errors; Lipoproteins; Clinical Chemistry Tests.