LINHARES, Isabela et al. Implementation of patient-specific quality control in radiotherapy treatments with ArcCheck. Rev. Fac. Med. Hum. [online]. 2024, vol.24, n.4, pp.43-51. Epub 31-Oct-2024. ISSN 1814-5469. http://dx.doi.org/10.25176/rfmh.v24i4.6793.
Introduction:
Quality control is essential to ensure safety and prevent errors in the administration of ionizing radiation across various radiotherapy techniques.
Objective:
To evaluate the performance of the ArcCheck detector to implement a specific quality control technique for patients treated with dynamic arc therapy.
Methods:
Fifty patients treated with 6 MV and 10 MV energies on the Clinac® Varian CX were selected. Doses at the isocenter of each treatment plan were analyzed using polymethyl methacrylate phantom (30 x 30 x 15 cm³) to validate reference values between the treatment planning system and the ionization chamber. The treatment plans were also recreated using the ArcCheck.
Results:
The mean dose difference at the isocenter was -0.96% and -1.34% for 6 MV and 10 MV, respectively. The average passing rate of the dose distributions in the gamma analysis exceeded 98.0% for both energies. Conclusion: The results demonstrated good concordance with the TG-119 and TG-218 protocols, supporting the use of the detector for quality control in patient-specific treatments.
Palabras clave : ArcCHECK; VMAT; Radiotherapy; gamma index. (Source: MESH-NLM).